| Title: Robust robot-camera calibration for robotized Transcranial Magnetic Stimulation. |
| Written by: L. Richter and F. Ernst and A. Schlaefer and A. Schweikard |
| in: <em>The International Journal of Medical Robotics and Computer Assisted Surgery</em>. (2011). |
| Volume: <strong>7</strong>. Number: |
| on pages: 414-422 |
| Chapter: |
| Editor: |
| Publisher: |
| Series: |
| Address: |
| Edition: |
| ISBN: |
| how published: |
| Organization: |
| School: |
| Institution: |
| Type: |
| DOI: 10.1002/rcs.411 |
| URL: |
| ARXIVID: |
| PMID: |
Note:
Abstract: Background: For robotized transcranial magnetic stimulation (TMS), the magnetic coil is placed on the patient\'s head by a robot. As the robotized TMS system requires tracking of head movements, robot and tracking camera need to be calibrated. However, for robotized TMS in a clinical setting, such calibration is required frequently. Mounting/unmounting a marker to the end effector and moving the robot into different poses is impractical. Moreover, if either system is moved during treatment, recalibration is required. Methods: To overcome this limitation, we propose to directly track a marker at link three of the articulated arm. Using forward kinematics and a constant marker transform to link three, the calibration can be performed instantly. Results: Our experimental results indicate an accuracy similar to standard hand-eye calibration approaches. It also outperforms classical hand-held navigated TMS systems. Conclusion: This robust online calibration greatly enhances the system\'s user-friendliness and safety