@inproceedings{RBSS10a,
Author = {L. Richter and R. Bruder and A. Schlaefer and A. Schweikard},
Title = {Towards direct head navigation for robot-guided transcranial magnetic stimulation using 3D laserscans: idea, setup and feasibility.},
Journal = {Conf Proc IEEE Eng Med Biol Soc.},
Year = {(2010).},
Pages = {2283-2286},
Booktitle = {Annual International Conference of the IEEE Engineering in Medicine and Biology Society},
Doi = {10.1109/IEMBS.2010.5627660},
Abstract = {Direct tracking is more robust than tracking that is based on additional markers. 3D laser scans can be used for direct tracking because they result in a 3D data set of surface points of the scanned object. For head\-navigated robotized systems, it is crucial to know where the patient\'s head is positioned relatively to the robot. We present a novel method to use a 3D laserscanner for direct head navigation in the robotized TMS system that places a coil on the patient\'s head using an industrial robot. First experimental results showed a translational error <= 2mm in the robot hand\-eye\-calibration with the laserscanner. The rotational error was 0.75\° and the scaling error <= 0.001. Furthermore, we found that the error of a scanned head to a reference head image was <= 0.2mm using ICP. These results have shown that a direct head navigation is feasible for the robotized TMS system. Additional effort has to be made in future systems to speed up the compution time for real time capability}
}