@conference{C09b,
author = {C. Fürweger and M. Kufeld and A. Schlaefer and C. Drexler},
title = {Fiducial-free spinal radiosurgery: Patient motion and targeting accuracy in 227 single fraction treatments with the Cyberknife.},
journal = {World Congress on Medical Physics and Biomedical Engineering IFMBE Proceedings.},
year = {2009},
volume = {25.},
number = {(1),},
pages = {277-280},
month = {September},
abstract = {Objective: To evaluate clinical targeting precision in fiducial\-free spinal treatments with a robotic radiosurgery system  (Cyberknife). Methods: For assessment of spine tracking system performance, we conducted phantom tracking tests on cervical and thoracic vertebrae. We retrospectively evaluated intrafraction patient movement for cervical (47), thoracic (90) and lumbar (90) treatments. A conservative measure for the expected targeting error due to patient motion was derived. Results: The phantom tests show that spinal targets are detected with an accuracy of \}
}